Reported Apr 8, 2026 · Ongoing
Brand Name: da Vinci S, Si Tenaculum Forceps Instrument Product Name: da Vinci S, Si Tenaculum Forceps Instrument…
Intuitive Surgical, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Intuitive Surgical, Inc. · 1266 Kifer Rd · Sunnyvale, CA · 94086-5304
Is this the product you have?
- Brand Nameda Vinci S, Si Tenaculum
- Forceps Instrument Product Nameda Vinci S, Si Tenaculum
- Forceps Instrument Model/Catalog Number420207
- Software VersionN/A
- Product DescriptionEndoWrist Instruments or Da Vinci S/Si Reusable instruments, including blunt and sharp endoscopic dissectors, scissors, scalpels, forceps/pick-ups, needle holders, endoscopic retractors, electrocautery and accessories are intended for endoscopic manipulation of tissue, including grasping, cutting, blunt and sharp dissection, approximation, ligation, electrocautery, suturing, and delivery and placement of microwave and cryogenic ablation probes and accessories. Please refer to attached User Manual for additional details on intended use. All da Vinci S/Si reusable instruments with jaws use mechanical tungsten cables and pulleys to transmit motion from the input discs to the housing (on the proximal chassis), through the main shaft, and the distal instrument tip (inside the patient)
- ComponentN/A
- Lot codes and dates
- Lot Code420207:
- Tenaculum Forceps Instrument GTIN00886874111659
- FDA Medical Device NumberD088920
- Material Number420207-06
- da Vinci System Numbers
SH1378SH0935SH0935SH0935USG294SH0761SH0729SH1182SH1378USG022SH1378SH1368SH0059SH1015SH1182SH0825USG294SH1390USG022SH0495SH0495SH1076SH1390SH1390SH1390SH1276SH1385USG022SH0547SH1385SH1391SH2144SH0059SH1279SH0059SH1390SH1570SH0059SH1391SG704SH1570SH0729SH1385SH1378USG518SH1385SH1837USG294SH1368SH1390SH1385SH1368SG2026SH1557SH2144SH1385USG518
- SH1279 USG294 SH1391 Material Number420207-07
- da Vinci System Numbers
100 of 100 codes
SH1391SH1385SH1385SH1279SH2125SH0729SH1368USG721SH1385SH1997SH2125USG294SH1378SH1997SH0035SG704SH1385SH1279SH1279SH2125SH1378SH1378SH0495SH1390SH1390SH1385SH2125SH1378SH1385SH1279USG721SH1279SH1837USG518SH1385SH1378USG022SH1378SH1570SH1385SH1385SH0935SH0935SG704SG704SH1378USG022SH1385SH1279USG721SH1279USG518SH1837SH1279USG022SH1385SH1385SH1964SH1385SH1390
- RSH0152 SH1378 SH0495 Material Number420207-09
- da Vinci System Numbers
RSH0152SH1378SH1385SH1279USG721SH1279SH1378SH1385USG721SH1279SH1182SH1390SH1076USG721SH1279SH1182SH1385SH1837
- SH1279 SH1837 SH1279 Material Number420207-10
- da Vinci System Numbers
SH1279SH1385SH1390SH1279SH1385SH1182SH1368SH1997SH1279USG721SH1822SH1557SH1822SH1964SH1279SH0495SH1390SH1279SH1964USG022SH1822SH1378SH1182SH1837SH1279SH1279SH1279SH1279SH1997SH1964SH1964SH1997SH0761SH1279SH1837SH1279SH1279SH1279SH1390SH1279SH1997
- Amount recalled
- 2095 instruments
This recall covers 6 products
The agency filed one notice per product. You are reading one of them; the other 5 are below.
Where it was sold
Why it was recalled
Due to increased in complaints regarding frayed or broken pitch cables on reusable surgical instruments
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Intuitive Surgical, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.