Reported Jun 11, 2014 · Terminated
Spacelabs Medical Ultraview SL Command Module, Model 91496. Arrhythmia Detector and Alarm. The Multi-parameter Module…
Spacelabs Healthcare Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Spacelabs Healthcare Inc · 35301 SE Center St · Snoqualmie, WA · 98065-9216
Is this the product you have?
- Spacelabs Medical Ultraview SL Command Module, Model 91496. Arrhythmia Detector and Alarm. The Multi-parameter Module is used with the Patient Care Management System (PCMS) to acquire, monitor, and process various clinical parameters from a patient in a clinical environment.
- Lot codes and dates
- All devices of Model91496
- Amount recalled
- 64 devices
Where it was sold
Why it was recalled
A software anomaly causes temperature alarm limits to change to limits which are not physiologically possible and result in nuisance false positive alarms.
This product could cause temporary or reversible health problems — stop using it.