Reported Apr 8, 2026 · Ongoing

MEDLINE ANTERIOR HIP PACK DYNJ64672B

Medline Industries, LP

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Medline Industries, LP · 3 Lakes Dr · Northfield, IL · 60093-2753

Is this the product you have?

  • MEDLINE ANTERIOR HIP PACK DYNJ64672B
Lot codes and dates
  • UDI-DI 10195327377496 (EA) 40195327377497 (CS) LOT 23LMD300
Amount recalled
14,379 kits total

This recall covers 10 products

The agency filed one notice per product. You are reading one of them; the other 9 are below.

Where it was sold

Why it was recalled

Kits contain recalled Olympus biopsy valves. Olympus is investigating an increase in complaints associated with rubber fragment detachment in the slit of single use biopsy valves. Potential consequences associated with detachment of fragments from valve include foreign body in patient's tracheobronchial tree, which may require intervention for removal. Other potential consequences include inflammatory response, hypoxia, and prolongation of procedure.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Medline Industries, LP recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 7:43 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.