Reported Jun 17, 2015 · Terminated
EVOTECH Endoscope Cleaner and Reprocessing System 208V, P/N 50004. The ASP EVOTECH ECR Endoscope Cleaner and…
Advanced Sterilization Products
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Advanced Sterilization Products · 33 Technology Dr · Irvine, CA · 92618-2346
Is this the product you have?
- EVOTECH Endoscope Cleaner and Reprocessing System 208V, P/N 50004. The ASP EVOTECH ECR Endoscope Cleaner and Reprocessor is a washer/disinfector, which is indicated for use with high-level disinfectant CIDEX OPA Concentrate and an enzymatic detergent (CIDEZYME Gl) to achieve cleaning and high level disinfection of heat sensitive (>60 ¿C) semi-critical endoscopes.
- Lot codes and dates
825 of 825 codes
504010706550401070735040107104504010710550401071065040107109504010711150401071125040107114504010711550401071215041070036504107003750410700385041070039504107004350410700445041070045504107004850410800135041080014504108002050410800575041080059504108006150410800625041080065504108006650410800685041080069504108007050410800715041080072504108007350410800745041080077504109002250410900235041090024504109002850410900315041090034504109005350410900545041090056504109005750410900585041090060504109006250410900645041090065504109006650410900675041090068504109006950410900705041090086504109008750410900905041090092
- Amount recalled
- 825 units
This recall covers 2 products
The agency filed one notice per product. You are reading one of them; the other one is below.
Where it was sold
- Nationwide
Why it was recalled
Advanced Sterilization Products (ASP) is recalling the EVOTECH Endoscope Cleaner and Reprocessor (ECR) because it may not detect an improper endoscope connection on two channels, and subsequently pass the cycle without cancellation if there is a disconnected scope.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Advanced Sterilization Products recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.