Reported Sep 21, 2022 · Ongoing

ZAP-X Radiosurgery System

Zap Surgical Systems

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Zap Surgical Systems · 590 Taylor Way · San Carlos, CA · 94070-6307

Is this the product you have?

  • ZAP-X Radiosurgery System
Lot codes and dates
  • UDI-DI
    • 00860183001504
    • Serial Numbers
    • ZUC17152
    • ZUC17153
    • ZUC18155
    • ZUC20157
    • ZUC20158
    • ZUC20156
    • ZUC21161
    • ZUC21162
    • using
  • Treatment Delivery Software Version1.8.53
Amount recalled
8

Where it was sold

Why it was recalled

Software issue identified in cases of initial patient setup with large (greater-than-or-equal-to 1.5 degrees) rotational deviations, between digitally reconstructed radiographs, and pre-delivery X-ray images. In these cases the transitional alignment algorithm may incorrectly calculate the new treatment table offset values, leading to an incorrect position for the subsequent treatment isocenter.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Zap Surgical Systems recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 5:44 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.