Reported May 28, 2025 · Ongoing
Bard¿ Blakemore Esophageal-Nasogastric Tube (Intermediate)REF: 0092300
C.R. Bard Inc.
High riskUsing this product could cause serious injury, illness or death — stop using it now.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: C.R. Bard Inc · 8195 Industrial Blvd Ne · Covington, GA · 30014-1497
Is this the product you have?
- Bard¿ Blakemore Esophageal-Nasogastric
- Tube (Intermediate)REF0092300
- Lot codes and dates
- All Lots within expiration/UDI(01)00801741076831
- Amount recalled
- N/A
This recall covers 5 products
The agency filed one notice per product. You are reading one of them; the other 4 are below.
Where it was sold
Why it was recalled
Stop using this product now
Due to customer complaints, there is a potential for users to be unable to or find it difficult to remove the plastic plugs in order to inflate the gastric and/or esophageal balloons.
Using this product could cause serious injury, illness or death — stop using it now.