Reported May 15, 2024 · Ongoing

Stealth S8 Clinical Software Application, REF: 9735762, CE0344, Rx Only. Used in the StealthStation System as an aid…

Medtronic Navigation, Inc.

High riskUsing this product could cause serious injury, illness or death — stop using it now.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Medtronic Navigation, Inc. · 200 Medtronic Drive · Lafayette, CO · 80026

Is this the product you have?

  • Stealth S8 Clinical Software Application
  • REF9735762, CE0344, Rx Only. Used in the StealthStation System as an aid for locating anatomical structures in either open or percutaneous neurosurgical procedures
Lot codes and dates
  • UDIs
    • 00763000234867
    • 00763000125295
    • 00763000063177
    • 00643169865341/
  • Software Versions
    • 1.2.0
    • 1.1.0
    • 1.0.3
    • 1.0.2
    • and 1.0.1
Amount recalled
3286 applications.

Where it was sold

  • Nationwide
  • PR

Why it was recalled

Stop using this product now

Due to a software issue, there is the potential for a missing digit, letter, or decimal in either the "distance to target" or "tip stop point" display during navigation.

Using this product could cause serious injury, illness or death — stop using it now.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Medtronic Navigation, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 5:40 PM UTC