Reported Jul 31, 2013 · Terminated

Philips HeartStart MRx Monitor/Defibrillator, Model numbers M3535A, M3536A, M3536J, M3536M, M3536MC, M3536M2, M3536M4,…

Philips Healthcare Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Philips Healthcare Inc. · 3000 Minuteman Road · Andover, MA · 01810

Is this the product you have?

  • Philips HeartStart MRx Monitor/Defibrillator, Model numbers M3535A, M3536A, M3536J, M3536M, M3536MC, M3536M2, M3536M4, M3536M5, M3536M6 The device is for use by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac support, or defibrillation. It must be used by or on the order of a physician
Lot codes and dates
  • Serial numbers ranging fromUS00100100 to US00567299 (serial numbers of the HeartStart MRx monitor/defibrillator are printed on the primary label on the back of the MRx in battery bay B)
Amount recalled
66,150 units

This recall covers 2 products

The agency filed one notice per product. You are reading one of them; the other one is below.

Where it was sold

  • Nationwide
  • PR

Why it was recalled

The MRx may deliver a non-synchronized cardioversion shock resulting in the delivery of incorrect therapy, which may induce ventricular fibrillation.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Philips Healthcare Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Dec 16, 2016 · Data as of Sep 19, 5:29 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.