Reported Apr 22, 2026 · Ongoing

B. Braun Hemodialysis Bloodlines, STREAMLINE BLOODLINE SET FOR DIALOG. Model Number: SL-2010M2096

B Braun Medical Inc.

High riskUsing this product could cause serious injury, illness or death — stop using it now.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: B Braun Medical Inc · 824 12th Ave · Bethlehem, PA · 18018-3524

Is this the product you have?

  • B. Braun Hemodialysis Bloodlines, STREAMLINE BLOODLINE SET FOR DIALOG
  • Model NumberSL-2010M2096
Lot codes and dates
  • Model NumberSL-2010M2096
  • UDI-DI Primary04046964367786
  • UDI-DI Unit
    • 04046964367779
    • All lots manufactured since 25MAY2025
Amount recalled
2,653,711 units

This recall covers 6 products

The agency filed one notice per product. You are reading one of them; the other 5 are below.

Where it was sold

  • Nationwide

Why it was recalled

Stop using this product now

Potential for the accumulation of small air bubbles in the arterial line due to adherence of blood gases to the tubing under negative pressure.

Using this product could cause serious injury, illness or death — stop using it now.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new B Braun Medical Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 4:02 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.