Reported May 6, 2020 · Completed

Autoimmune EIA Anti-Tissue Transglutaminase (tTG) IgA Test; Material Number 425-2260; Lot number 18200A - Product…

Bio-Rad Labs

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Bio-Rad Labs · 5500 E 2nd St · Benicia, CA · 94510-1020

Is this the product you have?

  • Autoimmune EIA Anti-Tissue Transglutaminase (tTG) IgA Test; Material Number 425-2260
  • Lot number 18200A - Product UsageThe assay is an aid in the diagnosis of celiac disease (gluten-sensitive enteropathy) and should be used in conjunction with other serological tests and clinical findings
Lot codes and dates
  • Catalog Number 425-2260
  • Lot Number 18200AUDI 00847817018336
Amount recalled
376 units

Where it was sold

Why it was recalled

An elevated lot-to-lot bias was observed for Autoimmune EIA Anti-Tissue Transglutaminase (tTG) IgA when compared to a previous lot of product. The observed elevated bias between these two lots is largely due to samples with higher levels of antibodies. With the cut-off for the assay being 15 U/mL, the clinical classifications for most samples would remain unchanged.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Bio-Rad Labs recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 6:59 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.