Reported Apr 22, 2026 · Ongoing
Yeastone Broth, 11ML, 10/BOX YY3462
Remel, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Remel, Inc · 12076 Santa Fe Trail Dr · Lenexa, KS · 66215-3519
Is this the product you have?
Yeastone Broth11ML10/BOX YY3462
- Lot codes and dates
UDI-DI 848838018817 lots 321046329835330316327069330317336325338314341161303851311723
- Amount recalled
- 2819
Where it was sold
- Nationwide
Why it was recalled
Products may report incorrect AST results during quality control. Health consequences may include delayed response or the need to change to another antifungal agent if laboratory ignores Quality Control test result.