Reported Jun 24, 2015 · Terminated
Philips Ultrasound, Model Q-Station, with software version 3 or higher, Catalog number: 795088; Part Number:…
Philips Ultrasound, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Philips Ultrasound, Inc. · 22100 Bothell Everett Hwy · Bothell, WA · 98021-8431
Is this the product you have?
- Philips Ultrasound, Model Q-Station, with software version 3 or higher
- Catalog number795088
- Part Number989605382391 Q-Station is application software intended to manage, view, analyze, and report qualitative and quantitative image data from ultrasound exams. It is designed to host optional advanced analysis applications via QLAB integration and provide integrated tools that allow users to manually assess and score cardiac wall motion and export images and / or exams and reports. Q-Station can view DICOM images of non-ultrasound images such as CT, MR, NM, CR, MG, XA, PET, RT and X-Ray modalities for reference viewing. It support connectivity to ultrasound systems, PACS and other DICOM storage repositories
- Lot codes and dates
- *****************US Serial numbers***************************************** 2176-IAF08E60-15D4, 2187-VAC5T6CD-157D, 2C98-25A23E3A-4B2E, 34F5-4084F543-80EA, 3A5D-4079F60B-80D4, 54B0-HADF24F2-15B1, 678B-NAE42F41-15BB, 6FF9-259B5EB7-4B20, 725B-GACE30D1-158F, 72EE-408914F8-80F2, 7A39-497627F7-92C9, 8536-IAD33B20-1599, A584-QAE84EF6-15C3, A838-IAE2CF5A-15B8, B903-407CB5D9-80D9, BDC7-TAE25A0D-15B7, D356-YAE1E4C0-15B6, D60A-GADC6524-15AB, D8BE-KAD6E588-15A0, DB72-RAD165EC-1595, DE26-WACBE650-158A, E0DA-XAC666B4-157F, E16D-40814ADB-80E2, E8E5-TAE16F73-15B5, EE4D-IAD6703B-159F, F669-PAC5F167-157E, F90C-MAEBF95E-15CA, H944-VACAFBB6-1589, MBF8-WAC57C1A-157E, U9D7-4085DFDD-80EC. ********************Serial numbers in CANADA**************************************** 1ED3-QACA8669-1588, 4489-SAD49B07-159C, B054-UAD25086-1597, C897-TACC5B9D-158B, and F3B5-IACB7103-1589. *********************Serial numbers in other countries*********************************** 114F-QAE6M475-15BF, 11F3-40756109-80CB, 16B7-KADBT53D-15A9, 196B-UAD585A1-159E, 1F66-40856A90-80EB, 221A-407FEAF4-80E0, 23FD-4977FD2B-92CD, 242A-MAEBPEC4-15C9, 24CE-407A6B58-80D5, 2782-4074EBBC-80CA, 2965-496CFDF3-92B7, 29E4-25A7BDD6-4B39, 2C46-QADA8FF0-15A8, 2EFA-RAD51054-159D, 2F4C-259CBE9E-4B23, 3716-IAC49180-157C, 37A9-407F75A7-80DF, 3C6D-XAE519DB-15BD, 3CBF-25ACC825-4B43, 3D11-4074766F-80C9, 41D5-NADA1AA3-15A7, 47D0-4089FF92-80F4, 49F1-IAC99BCF-1586, 4C94-YAEFA3C6-15D2, 4CA5-UAC41C33-157B, 4F1B-49771291-92CB, 4FEC-407980BE-80D3, 524E-25AC52D8-4B42, 5764-MAD9A556-15A6, 5A18-QAD425BA-159B, 5CCC-HACEA61E-1590, 5F80-MAC92682-1585, 6013-4084IAA9-80E8, 61F6-497C1CE0-92D5, 62C7-407E8B0D-80DD, 657B-4079TB71-80D2, 6CF3-WAD93009-15A5, 750F-TAC8B135-1584, 75A2-4083955C-80E7, 7856-407E15C0-80DC, 7B0A-40789624-80D1, 7CED-4970A85B-92BE, 7D6C-25AB683E-4B40, 7FCE-HADE3A58-15AF, 8282-JAD8BABC-15A4, 87EA-PACDBB84-158E, 8A9E-RAC83BE8-1583, 8B31-4083200F-80E6, 8D14-497B3246-92D3, 8D41-NAEE43DF-15CF, 8DE5-407DA073-80DB, 8FF5-NAE8C443-15C4, 9099-407820D7-80D0, 9811-WAD8456F-15A3, 9D79-WACD4637-158D, A0C0-4082AAC2-80E5, A374-407D2B26-80DA, A628-4077AB8A-80CF, A80B-496FBDC1-92BC, AAEC-MADD4FBE-15AD, ADA0-PAD7D022-15A2, B308-JACCD0EA-158C, B39B-4087B511-80EF, B5BC-RAC7514E-1581, B64F-40823575-80E4, B832-497A47AC-92D1, B85F-UAED5945-15CD, BAE6-4974C810-92C6, BBB7-4077363D-80CE, BE19-25AAH857-4B3D, C07B-KADCDA71-15AC, CBDE-4081C028-80E3, CDC1-4979D25F-92D0, CE92-407C408C-80D8, D146-4076C0F0-80CD, E37D-JAEC6EAB-15CB, E421-407BCB3F-80D7, F448-4086552A-80EC, F6FC-4080D58E-80E1, F8DF-4978E7C5-92CE, F9B0-407B55F2-80D6, FC64-4075D656-80CB, T3DC-SAD5FAEE-159F, TF3F-407AE0A5-80D6, VC8B-40806041-80E1, and Y690-GAD07B52-1594. **********NEWLY ADDED FOUR AFFECTED SERIAL NUMBERS******* aaec-madd4fbe-15ad; WE9B-QAEB8411-15CA; 2B14-MDE9A2F9-1BC6; and 3C40-4972P842-92C1
- Amount recalled
- 125 units (30 in US and 95 international)
Where it was sold
- Nationwide
- PR
Why it was recalled
Philips Healthcare has discovered a problem in the Philips Ultrasound Q-Station version 3.0 or higher that could result in measurements from a study (Structured Report) to be appended to subsequent studies for other patients.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Philips Ultrasound, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.