Reported May 6, 2020 · Terminated

Philips Azurion 7 M20 -XperGuide Software hosted in Interventional Workspot 1.5 when used with the Azurion R2.0 system…

Philips North America, LLC

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Philips North America, LLC · 3000 Minuteman Rd · Andover, MA · 01810-1032

Is this the product you have?

  • Philips Azurion 7 M20 -XperGuide Software hosted in Interventional Workspot 1.5 when used with the Azurion R2.0 system Interventional
  • fluoroscopic x-ray system System code722079
Lot codes and dates
  • SoftwarePhilips Azurion 2.0 Interventional Fluoroscopic X-ray system with Interventional Workspot 1.5
  • US Serial Numbers
    • 637
    • 200
    • 909
    • 1276
    • 1309
    • 987
  • Canada1016
  • ROW
    • 11169
    • 1129
    • 1143
    • 971
    • 1135
    • 824
    • 934
    • 1089
    • 1008
    • 1165
    • 1080
    • 1019
    • 1248
    • 523
    • 1172
  • Expanded Recall OUS
    • 1223
    • 1330
    • 703424
Amount recalled
22 units (6 US and 16 OUS); Expanded 3units; Total: 25

This recall covers 3 products

The agency filed one notice per product. You are reading one of them; the other 2 are below.

Where it was sold

Why it was recalled

When a user acquires XperCT scan on an Azurion 2.0 system, enters the XperGuide guidance step and moves the L-arm away from the initial scan position before starting the live guidance, a warning message directs the user to move the L-arm stand back to the initial XperCT scan position. Although the software generates this message, it does not prevent the use of live guidance if the L-arm stand is not repositioned. Using live guidance with a mispositioned L-arm can result in the display of an incorrect overlay and needle path.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Philips North America, LLC recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Feb 1, 2023 · Data as of Sep 19, 5:26 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.