Reported Jun 24, 2015 · Terminated
Thoracic Grasper, 5 MM, for use with da Vinci S (IS2000) and SI (IS3000) Surgical Systems. Intended for endoscopic…
Intuitive Surgical, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Intuitive Surgical, Inc. · 1266 Kifer Rd Bldg 100 · Sunnyvale, CA · 94086-5304
Is this the product you have?
- Thoracic Grasper, 5 MM, for use with da Vinci S (IS2000) and SI (IS3000) Surgical Systems. Intended for endoscopic manipulation of tissue, including grasping.
- Lot codes and dates
- version 420343-01 and 420323-02, all lots
- Amount recalled
- 2,436 units
Why it was recalled
The coating of the 5 mm Thoracic Grasper main tube may experience damage with the possibility of fragments of coating material dislodging and potentially falling into patient.
This product could cause temporary or reversible health problems — stop using it.