Reported Oct 5, 2022 · Ongoing

Olympus POWERSEAL 5MM, 37CM, CURVED JAW SEALER & DIVIDER, DOUBLE-ACTION Model Number: PS-0537CJDA. An electrosurgical…

Olympus Corporation of the Americas

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Olympus Corporation of the Americas · 3500 Corporate Pkwy · Center Valley, PA · 18034-8229

Is this the product you have?

    • Olympus POWERSEAL 5MM
    • 37CM
    • CURVED JAW SEALER & DIVIDER
  • DOUBLE-ACTION Model NumberPS-0537CJDA. An electrosurgical device use in laparoscopic/minimally invasive or open surgical procedures
Lot codes and dates
  • UDI-DI00821925044555
  • Lot Numbers
    • CA191233
    • CA191234
    • CA191235
    • CA233514
    • CA233515
    • CA233516
    • CA233517
    • CA233518
    • CA236206
    • CA236208
    • CA236213
    • CA236215
    • CA236216
    • CA236217
    • CA236220
    • CA236227
    • CA236228
    • CA236230
    • CA236238
    • CA236240
    • CA236243
    • CA236261
    • CA236272
    • CA236273
    • CA236276
    • CA236278
    • CA236290
Amount recalled
231.4 units (5/box); 510 units (5/box) OUS

This recall covers 3 products

The agency filed one notice per product. You are reading one of them; the other 2 are below.

Where it was sold

  • Nationwide

Why it was recalled

An out of specification component preventing the blue retention ring from keeping the rotation knob in place. This may result in a loose rotation collar and retention ring and cause delay of treatment

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Olympus Corporation of the Americas recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 5:42 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.