Reported Jun 4, 2025 · Ongoing

Apo B Reagent, REF: OSR6143

Beckman Coulter Ireland, Inc.

Low riskThis product is unlikely to cause harm, but it breaks a safety or labelling rule.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Beckman Coulter Ireland, Inc. · Lismeehan · O'Callaghan'S Mills, N/A · N/A

Is this the product you have?

  • Apo B Reagent
  • REFOSR6143
Lot codes and dates
  • Lot # 2631/UDI: 15099590010409
Amount recalled
2747 units

Where it was sold

Why it was recalled

The product Value Assignment Sheet (VAS) is used in calibration of the device. An erroneous value was included in the VAS which has the potential to generate an incorrect APO B patient result which could exceed the total allowable error for the APO B assay (11.6%). The error results in results with a positive bias up to 42%. However, based on the mitigating factors the risk to health is considered to be negligible or limited.

This product is unlikely to cause harm, but it breaks a safety or labelling rule.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Beckman Coulter Ireland, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 4:50 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.