Reported Jul 1, 2015 · Terminated
ORTHOFIX Centurion POCT System Drill Guide PN 69-1010 and PN 69-1012. Drill Guides are used to interface with various…
Orthofix, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Orthofix, Inc · 3451 Plano Pkwy · Lewisville, TX · 75056-9453
Is this the product you have?
- ORTHOFIX Centurion POCT System Drill Guide PN 69-1010 and PN 69-1012. Drill Guides are used to interface with various drill tips during the drilling procedure of bone preparation.
- Lot codes and dates
- All lot codes
- Amount recalled
- 36 units
Where it was sold
Why it was recalled
There is a possibility that the thumb button remains in the depressed (unlocked) position allowing the threaded sleeve (drill stop) to unintentionally rotate and change the depth setting.
This product could cause temporary or reversible health problems — stop using it.