Reported Jun 27, 2012 · Terminated

Philips Easy Diagnost Product Usage: Stationary Fluoroscopic system

Philips Healthcare Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Philips Healthcare Inc. · 3000 Minuteman Road · Andover, MA · 01810

Is this the product you have?

  • Philips Easy Diagnost Product UsageStationary Fluoroscopic system
Lot codes and dates
  • All systems with Eleva software version 4.x and stitching option.
Amount recalled
6

This recall covers 3 products

The agency filed one notice per product. You are reading one of them; the other 2 are below.

Why it was recalled

If a lead ruler is not positioned in the same plane as the anatomical image detail to be measured, measured object distances (using the ruler) can be wrong by a corresponding factor. If, additionally, the measurement tool is not used to check the ruler, or is used in the right way (ie: calibration), such calibration error is not detected. As a result of this, a surgical intervention may be execut

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Philips Healthcare Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Oct 6, 2016 · Data as of Sep 19, 5:26 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.