Reported Jul 1, 2015 · Terminated

Surface Applicator Set with Leipzig-style Cone. Brachytherapy applicator set.

Varian Medical Systems Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Varian Medical Systems Inc · 501 Locust Ave · Charlottesville, VA · 22902-4869

Is this the product you have?

  • Surface Applicator Set with Leipzig-style Cone. Brachytherapy applicator set.
Lot codes and dates
  • JAQ
Amount recalled
46

Why it was recalled

Varian has discovered that there is a discrepancy in the absolute dose rate given with the Leipzig-style Surface Applicator's Instructions for Use (IFU): Dose Characterization GM11010080 2012-09-06. The actual dose rate of the applicator is approximately 14% higher than the rate published within the IFU. Use of the dose rate as provided by the IFU without modification or independent confirmat

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Varian Medical Systems Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Jan 26, 2016 · Data as of Sep 19, 6:15 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.