Reported Aug 7, 2013 · Terminated

AIA Analyzer Pipette Tips used on Tosoh AIA-2000, AIA-1800, AIA-900 and AIA-600II Analyzers, 1000 tips/bag AIA-Analyzer…

Tosoh Smd Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Tosoh Smd Inc · 3600 Gantz Rd · Grove City, OH · 43123-1895

Is this the product you have?

  • AIA Analyzer Pipette Tips used on Tosoh AIA-2000, AIA-1800, AIA-900 and AIA-600II Analyzers, 1000 tips/bag AIA-Analyzer Tips are disposable pipette tips packaged in bulk quantities and used on Tosoh Bioscience AIA Analyzers for lab analysis of various immunoassays.
Lot codes and dates
  • The following lots
  • numbers are affected by this recall
    • 060811
    • 070111
    • B11
    • B11-12/23/2011
    • B12-02/07/2012
    • B12-04/30/2012
    • B12-06/01/2012
    • B12-08/06/2012
    • B12-09/05/2012
    • B12-10/26/2012
    • and B12-11/27/2012
Amount recalled
11,309 bags

Where it was sold

Why it was recalled

Visual inspection of the pipette tips indicated that some of the tips were not uniform in size/length which was causing error messages on the analyzers.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Tosoh Smd Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Nov 27, 2013 · Data as of Sep 19, 4:50 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.