Reported Oct 5, 2022 · Ongoing
Rosen Curved Wire Guide, Guidewire Reference Part Number/GPN THSCF-35-220-1.5-ROSEN G01623 THSCF-35-80-1.5-ROSEN G01774…
Cook Incorporated
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Cook Incorporated · 750 N Daniels Way · Bloomington, IN · 47404-9120
Is this the product you have?
4 of 4 products
| Model / number | Product as published |
|---|---|
G01623 | Rosen Curved Wire Guide, Guidewire Reference Part Number/GPN THSCF-35-220-1.5-ROSEN |
G01774 | THSCF-35-80-1.5-ROSEN |
G01264 | THSCF-35-180-1.5-ROSEN |
G01261 | THSCF-35-145-1.5-ROSEN |
- Lot codes and dates
4 of 4 products
Products covered by this recall # Barcodes Product as published 1 00827002017740Lots/Expiration Date: 14818344 29-06-2027 14820965 30-06-2027 14849857 20-07-2027 THSCF-35-80-1.5-ROSEN G01774 UDI-DI: 00827002017740 2 00827002012646Lots/Expiration Date: 14820960 30-06-2027 14832877 08-07-2027 14847945 19-07-2027 THSCF-35-180-1.5-ROSEN G01264 UDI-DI: 00827002012646 3 00827002012615Lots/Expiration Date: 14820961 30-06-2027 14843133 15-07-2027 14847946 19-07-2027 THSCF-35-145-1.5-ROSEN G01261 UDI-DI: 00827002012615 4 Lots/Expiration Date: 14820962 30-06-2027 14820963 30-06-2027 14820964 30-06-2027 14828587 06-07-2027 14835364 11-07-2027 14835365 11-07-2027 14849859 20-07-2027 - Amount recalled
- 5041 US; 298 OUS
This recall covers 24 products
The agency filed one notice per product. You are reading one of them; the other 23 are below.
11 more are in the archive.
Where it was sold
- Nationwide
Why it was recalled
Cook Medical identified that devices from the affected device lots may have a complete breach of the chevron seal of the packaging. Therefore, the sterility of affected devices may be compromised.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Cook Incorporated recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.