Reported Jun 8, 2016 · Terminated
EZ Slide Battery/line powered hydraulic table with removable leg section and 21 top slide General Purpose Surgical…
Skytron, Div. The KMW Group, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Skytron, Div. The KMW Group, Inc · 5085 Corporate Exchange Blvd SE · Grand Rapids, MI · 49512-5515
Is this the product you have?
- EZ Slide Battery/line powered hydraulic table with removable leg section and 21 top slide General Purpose Surgical Table Skytron EZ Slide Operating Table General Purpose Surgical Table
- Lot codes and dates
- Model number 3502. All units. Units entered commercial distribution between April 30, 2014 and July 24, 2014
- Amount recalled
- 458
This recall covers 2 products
The agency filed one notice per product. You are reading one of them; the other one is below.
Where it was sold
Why it was recalled
The potential exists within the identified tables that incorrectly manufactured parts have been used in assembly. If a table is found to have these incorrectly manufactured parts, it is possible for the tables elevation column slide cap to become detached from the elevation column and could result in the loss of control of the table top and injury to the patient or staff.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Skytron, Div. The KMW Group, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.