Reported Jul 4, 2012 · Terminated

Go Bed II+ Model FL28C Beds are intended for medical purposes in both general patient care as well as critical care…

Stryker Medical Division of Stryker Corporation

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Stryker Medical Division of Stryker Corporation · 3800 East Centre Ave. · Portage, MI · 49002

Is this the product you have?

  • Go Bed II+ Model FL28C Beds are intended for medical purposes in both general patient care as well as critical care environments. Each uses built-in electric motors and controls which can be operated by the patient, or caregiver, to adjust the height and/or surface contour of the bed.The devices listed above also include movable and latchable side rails which can be lowered or raised to help prevent patient fall and subsequent injury. The
  • Beds include an optional Scale (ModelsFL20E, FL25E, FL28C, FL28EX) as well as an optional Bed Exit (1 Zone) or Chaperone
Lot codes and dates
  • Model FL28C

    4,072 of 4,072 codes

    • Serial Number R25210
    • R21286
    • R20899
    • R16969
    • R16968
    • R15495
    • R15494
    • R15493
    • R15492
    • R15491
    • R15490
    • R15489
    • R15488
    • R15487
    • R15486
    • R15485
    • R15484
    • R15483
    • R15482
    • R15481
    • R15480
    • R15479
    • R15478
    • R15447
    • R15446
    • R15445
    • R15444
    • R15443
    • R15442
    • R15441
    • R15440
    • R15439
    • R15438
    • R15437
    • R15436
    • R15435
    • R15434
    • R15433
    • R15432
    • R15431
    • R15430
    • R15429
    • R15428
    • R15427
    • R15426
    • R15425
    • R15424
    • R15423
    • R15422
    • R15421
    • R15420
    • R15419
    • R15418
    • R15417
    • R15132
    • R15131
    • R15130
    • R15129
    • R15128
    • R15127
Amount recalled
8009

This recall covers 5 products

The agency filed one notice per product. You are reading one of them; the other 4 are below.

Where it was sold

  • Nationwide
  • PR

Why it was recalled

An adverse trend of scale and bed exit service reports/complaints were detected for the GoBed+, GoBed II, Rose, and MA204 model beds. Investigation found an abnormally high failure rate of loadcells, an integral component in scale and bed exit function. The scale faults are manifested as a fluctuating scale reading, a scale error causing the scale system to be disabled until serviced, or an inac

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Stryker Medical Division recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated May 12, 2014 · Data as of Sep 19, 5:25 PM UTC