Reported Jul 3, 2019 · Terminated
Fisher Scientific Isotemp -General Purpose Water Bath: lsotemp 10L GP Bath Model Number: FSGPD10
Thermo Fisher Scientific (Asheville) LLC
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Thermo Fisher Scientific (Asheville) LLC · 401 Mill Creek Rd · Marietta, OH · 45750-4304
Is this the product you have?
- Fisher Scientific
- Isotemp -General Purpose Water Bathlsotemp 10L
- GP Bath Model NumberFSGPD10
- Lot codes and dates
- Serial Numbers300247769 300247772 300247773 300247774 30024777S
- Amount recalled
- 5 units
This recall covers 9 products
The agency filed one notice per product. You are reading one of them; the other 8 are below.
Where it was sold
Why it was recalled
Product may have been assembled incorrectly,the assembly error can cause the bath to malfunction and raise the temperature of the fluid in the reservoir above the temperature set point selected by the user.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Thermo Fisher Scientific (Asheville) LLC recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.