Reported May 3, 2017 · Terminated

01.4mm x 457mm (18in) K-Wire, Threaded Blunt, Ref 9080B-18T, Qty: 1 per pack, Material: SS 316L, Full Temper, per ASTM…

Amendia, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Amendia, Inc · 1755 W Oak Pkwy · Marietta, GA · 30062-2260

Is this the product you have?

    • 01.4mm x 457mm (18in) K-Wire
    • Threaded Blunt
    • Ref 9080B-18T
  • Qty1 per pack
  • MaterialSS 316L, Full Temper, per ASTM F138, Amendia, 1755 West Oak Parkway, Marietta, GA 30062 Guide Wires are used to establish a pathway into the bony pedicle to place implants during spinal surgery
Lot codes and dates
  • Lot Number 135561
Amount recalled
15 units

Where it was sold

Why it was recalled

Mislabeled guide wires - It was discovered that some Guide Wires labeled as stainless steel were actually manufactured from nitinol.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Amendia, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated May 30, 2017 · Data as of Sep 19, 4:45 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.