Reported Jun 22, 2016 · Terminated
AVEA Comprehensive ventilator, 17310-xx (xx = 0 - 14) A continuous ventilator intended to mechanically control or…
Carefusion 211 Inc. dba Carefusion
High riskUsing this product could cause serious injury, illness or death — stop using it now.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Carefusion 211 Inc dba Carefusion · 22745 Savi Ranch Pkwy · Yorba Linda, CA · 92887-4668
Is this the product you have?
- AVEA Comprehensive ventilator, 17310-xx (xx = 0 - 14) A continuous ventilator intended to mechanically control or assist broad spectrum patient breathing by delivering a predetermined percentage of oxygen in the breathing gas.
- Amount recalled
- 922 units total (501 units in US)
This recall covers 12 products
The agency filed one notice per product. You are reading one of them; the other 11 are below.
Where it was sold
- Nationwide
Why it was recalled
Stop using this product now
CareFusion has identified a potential risk associated with AVEA Ventilator caused by an incorrect F1 fuse on the AVEA Ventilator Transducer Communication Alarm (TCA) Board which may fail/blow, causing loss of power to the User Interface Module (UIM).
Using this product could cause serious injury, illness or death — stop using it now.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Carefusion 211 Inc. dba Carefusion recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.