Reported Jun 11, 2025 · Ongoing

Signature Laparoscopic Instruments, Debakey Grasper with Rotating Shaft and Monopolar Electrosurgical Footswitch…

Silex Medical, LLC

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Silex Medical, LLC · 24 Robert Porter Rd Ste B · Southington, CT · 06489-1158

Is this the product you have?

  • Signature Laparoscopic Instruments, Debakey Grasper with Rotating Shaft and Monopolar Electrosurgical Footswitch Connector
  • Model/REF Number5862100
Lot codes and dates
  • Model/REF Number5862100
  • UDI-DIB33158621000
  • Lot NumberM28183
  • Serial Numbers
    • 1932
    • 1933
    • 1934
    • 1935
    • 1936
    • 1937
    • 1938
    • 1939
Amount recalled
8 units

This recall covers 8 products

The agency filed one notice per product. You are reading one of them; the other 7 are below.

Where it was sold

Why it was recalled

Insulation on the tube did not extend fully to the proximal end of the shaft on the instrument, leaving an exposed (not insulated) area of the shaft closer to the rotator wheel.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Silex Medical, LLC recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 7:35 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.