Reported Jun 23, 2021 · Terminated
OSSUR MIAMI J SELECT COLLAR - Product Usage: is intended to provide gross immobilization to the cervical spine. REF:…
Ossur Americas
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Ossur Americas · 27051 Towne Centre Dr · Foothill Ranch, CA · 92610-2804
Is this the product you have?
- OSSUR
- MIAMI J SELECT COLLAR - Product Usageis intended to provide gross immobilization to the cervical spine
- REF
MJS-101MJSR-101
- Lot codes and dates
- REF Number
MJS-101MJSR-101
- Lot NumbersMX180425 through MX200306
- REF Number
- Amount recalled
- 10,469 items
Where it was sold
Why it was recalled
Due to insufficient column strength of the racks for all load conditions, racks associated with height adjustment functionality have been observed deforming, which may lead to increased range of motion in the cervical spine.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Ossur Americas recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.