Reported Jun 6, 2018 · Ongoing
Heartware Medtronic HVAD System for cardiac use. Including the following parts: (a) Controller / Controller Kits,…
Heartware
High riskUsing this product could cause serious injury, illness or death — stop using it now.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Heartware · 14400 NW 60th Ave · Miami Lakes, FL · 33014-2807
Is this the product you have?
- Heartware Medtronic HVAD System for cardiac use
- Including the following parts(a) Controller / Controller Kits
- Product numbers
14001401140314071420 (b) DC Adapter
- Product numbers
14351440 (c) AC Adapter
- Product numbers
14251430 (d) Battery Pack
- Product number1650DE
- Indication for Use for OUS HVAD SystemThe HVAD System is intended for use in patients at risk of death from refractory end-stage heart failure. The HVAD System is designed for in-hospital and out-of-hospital settings, including transportation via fixed wing aircraft or helicopter
- Lot codes and dates
- ALL SERIAL NUMBERS UPN/GTIN(a) 1400
- and 1401distributed
- prior to GTN requirement 140300888707000116, 00888707000475 1407: '00888707000291, '00888707000727, '00888707000710, 00888707000734, '00888707000741, '00888707000765, '00888707000758, '00888707000789, '00888707000772, '00888707000499, '00888707000796, '00888707000802, '00888707000819, '00888707000482, '00888707004534, '0888707001670, '00888707001656, '00888707001663, '00888707001670, '00888707001687, '00888707001694, '00888707001700, '00888707001717, '00888707001724, '00888707001731, '00888707001748, '00888707001755, '00888707001762, '00888707002813, '00888707002820, '00888707002837, '00888707002844, '00888707002851, '00888707001472 1420: '00888707000420, '00888707000437, '00888707002660 (b) 1435: '00888707000109 1440: '00888707000260, '00888707001885, '00888707001496, '00888707002745 (c) 1425: '00888707000093 1430: '00888707000307, '00888707000826, '00888707000833, '00888707000840, '00888707000857, '00888707000871, '00888707000864, '00888707000888, '00888707000901, '00888707000505, '00888707000895, '00888707000918, '00888707001489, '00888707001779, '00888707001786, '00888707001793, '00888707001809, '00888707001816, '00888707001823, '00888707001830, '00888707001847, '00888707001854, '00888707001861, '00888707001878, '00888707002738, '00888707002769, '00888707002776, '00888707002783, '00888707002790, '00888707002806, '00888707004817 (d) 1650DE: 00888707000369, 00888707001373, '00888707001588, '00888707001366, '00888707002684 '00888707002646, '00888707002653, '00888707000376
- Amount recalled
- 175878 devices
Where it was sold
- Nationwide
Why it was recalled
Stop using this product now
Possible transient electrical connection interruption between an HVAD System power source (Battery, AC Adapter, or DC Adapter) and the Controller, while the power source is connected, that may cause unintended switching to the secondary power source and/or unexpected audible tones (beeping). The interruption typically lasts 1-2 seconds, due to oxidation between power source connector and socket.
Using this product could cause serious injury, illness or death — stop using it now.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Heartware recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.