Reported May 13, 2020 · Terminated

REF: Z2850, 112" (284 cm) 10 Drop Primary set w/2 Clave, Remv 3 Gang 1o2 Manifold w/3 Clave, Baseplate, Rotating Luer…

ICU Medical, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: ICU Medical, Inc. · 951 Calle Amanecer · San Clemente, CA · 92673-6212

Is this the product you have?

  • REFZ2850, 112" (284 cm) 10 Drop Primary set w/2 Clave, Remv 3 Gang 1o2 Manifold w/3 Clave, Baseplate, Rotating Luer w/Filter Cap, 1 Ext, Sterile R
  • UDI(01)00887709028241
Lot codes and dates
  • Lot number4172062
Amount recalled
25 units

This recall covers 11 products

The agency filed one notice per product. You are reading one of them; the other 10 are below.

Where it was sold

Why it was recalled

There is a potential that IV administration sets contain a 15 drops/mL drop cannula instead of a 10 drops/mL drop cannula within the drip chamber. An infusion set with an incorrect drop cannula could result in inaccurate delivery of the intended infusion volume (under-delivery). This could result in under-delivery of critical medication which may lead to serious adverse health consequences.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new ICU Medical, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Dec 28, 2021 · Data as of Sep 19, 6:51 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.