Reported May 24, 2017 · Terminated

Magellan Diagnostics LeadCare Ultra Blood Lead Test Kits Part Number: 80-0010

Magellan Diagnostics, Inc.

High riskUsing this product could cause serious injury, illness or death — stop using it now.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Magellan Diagnostics, Inc. · 101 Billerica Ave Bldg 4 · North Billerica, MA · 01862-1271

Is this the product you have?

  • Magellan Diagnostics LeadCare
  • Ultra Blood Lead Test Kits Part Number80-0010
Lot codes and dates
  • Kit lots
    • 1309BU
    • 1310AU
    • 1312BU
    • 1404AU
    • 1408BU
Amount recalled
1016

This recall covers 2 products

The agency filed one notice per product. You are reading one of them; the other one is below.

Where it was sold

  • Nationwide

Why it was recalled

Stop using this product now

LeadCare Plus and LeadCare Ultra Blood Lead Testing Systems may generate falsely low lead concentrations in some blood samples. The issue is observed when blood samples are collected in tubes with lavender and tan rubber stoppers of the Becton Dickinson EDTA Vacutainer tubes, a collection device used with the testing system.

Using this product could cause serious injury, illness or death — stop using it now.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Magellan Diagnostics, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Feb 26, 2022 · Data as of Sep 19, 6:15 PM UTC