Reported Jul 8, 2015 · Terminated
LDL-Cholesterol gen.2. The Boehringer Mannheim Direct LDL-Cholesterol test in intended for the quantitative…
Roche Diagnostics Operations, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Roche Diagnostics Operations, Inc. · 9115 Hague Rd · Indianapolis, IN · 46256-1025
Is this the product you have?
- LDL-Cholesterol gen.2. The Boehringer Mannheim Direct LDL-Cholesterol test in intended for the quantitative determination of low-density lipoprotein Cholesterol (LDL-C) in serum and plasma.
- Lot codes and dates
Catalog numbers 04714423190047112201900523043819005230446190
- Amount recalled
- 6255
This recall covers 12 products
The agency filed one notice per product. You are reading one of them; the other 11 are below.
Where it was sold
- Nationwide
- PR
Why it was recalled
Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of Trinder tests may be falsely decreased when the tests are run with urine samples (Creatinine plus and Uric Acid applications only) or blood samples collected before sufficient time has lapsed after the administration of the drugs Acetaminophen and N-acetylcysteine.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Roche Diagnostics Operations, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.