Biograph Horizon, PET/CT System using VJ10A, VJ10B, VJ20A scanners that provide registration and fusion of high…
Siemens Medical Solutions USA, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Siemens Medical Solutions USA, Inc. · 2501 Barrington Rd · Hoffman Estates, IL · 60192-2061
Is this the product you have?
- Biograph Horizon, PET/CT System using VJ10A, VJ10B, VJ20A scanners that provide registration and fusion of high resolution physiologic and anatomic information.
- Lot codes and dates
Code No. 1053274610532748
- Amount recalled
- 14 units
Why it was recalled
Possibility that the Biograph Horizon systems performing CT retrospective cardiac gating or PET cardiac gating examinations may experience a waveform sampling issue caused by a firmware change within the Universal Physiological Measurement Module.