Reported Jul 10, 2019 · Terminated
RayStation 3.5, RayStation 4.0, RayStation 4.5, RayStation 4.7, RayStation 5, RayStation 6, RayPlan 1 (RayStation 4.9)…
Raysearch Laboratories Ab
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: RAYSEARCH LABORATORIES AB · Sveavagen 9 · Stockholm, N/A · N/A
Is this the product you have?
- RayStation 3.5, RayStation 4.0, RayStation 4.5, RayStation 4.7, RayStation 5, RayStation 6, RayPlan 1 (RayStation 4.9) and RayPlan 2 (RayStation 6)
- Intended Usesoftware system designed for treatment planning and analysis of radiation therapy
- Lot codes and dates
- Software versions
3.5.0.163.5.1.64.0.0.144.0.1.44.0.2.94.0.3.44.5.0.194.5.1.144.5.2.74.7.0.154.7.1.104.7.2.54.7.3.134.7.4.44.7.5.44.9.0.425.0.0.375.0.1.115.0.2.355.0.3.176.0.0.246.1.0.266.1.1.2 and 6.2.0.7
- Software versions
- Amount recalled
- 744
Where it was sold
- Nationwide
Why it was recalled
When calculating electron Monte Carlo dose with a very large number of histories, the dose calculation may be wrong
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Raysearch Laboratories Ab recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.