Reported Jun 15, 2016 · Terminated

Maestro Air Motor Pneumatic cranial drill motor. The Maestro Air Pneumatic System is a pneumatically operated surgical…

Stryker Instruments Div. of Stryker Corporation

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Stryker Instruments Div. of Stryker Corporation · 4100 E Milham Ave · Portage, MI · 49002-9704

Is this the product you have?

  • Maestro Air Motor Pneumatic cranial drill motor. The Maestro Air Pneumatic System is a pneumatically operated surgical instrument system. The pneumatic motor provides power to operate removable rotating surgical cutting tools and their accessories.
Lot codes and dates
  • Model Number5407-200-000
  • S/N

    189 of 189 codes

    • 1522406703
    • 1522406713
    • 1522406733
    • 1522406743
    • 1522406753
    • 1522406763
    • 1522406773
    • 1522406793
    • 1522406803
    • 1522406813
    • 1522406833
    • 1522406843
    • 1522406853
    • 1522406863
    • 1522406873
    • 1522406883
    • 1522406893
    • 1522406903
    • 1522406923
    • 1523018583
    • 1523018603
    • 1523018623
    • 1523018633
    • 1523018683
    • 1523018713
    • 1523018723
    • 1523018733
    • 1523018743
    • 1523018773
    • 1523018783
    • 1523018813
    • 1523018823
    • 1523018833
    • 1523018843
    • 1523018853
    • 1523018863
    • 1523018873
    • 1523018883
    • 1523018893
    • 1523018903
    • 1523018913
    • 1523601703
    • 1523601723
    • 1523601733
    • 1523601743
    • 1523601763
    • 1523601773
    • 1523601783
    • 1523601793
    • 1523601813
    • 1523601823
    • 1523601833
    • 1523601843
    • 1523601863
    • 1523601873
    • 1523601883
    • 1523601893
    • 1523601903
    • 1523601913
    • 1523601923
Amount recalled
189

Where it was sold

Why it was recalled

Due to the buildup of grease within the hose and motor, the product may require multiple attempts to start. The grease may also migrate into the attachment end of the motor. There is a negligible possibility that the orange grease may leak from the attachment end of the motor and contaminate the surgical wound resulting in soft tissue inflammation.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Stryker Instruments Div. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Mar 24, 2017 · Data as of Sep 19, 6:46 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.