Reported May 6, 2026 · Ongoing

AIDBAGs are first aid kits of convenience composed of individually labeled devices designed as specialty first aid kits…

North American Rescue LLC.

High riskUsing this product could cause serious injury, illness or death — stop using it now.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: North American Rescue LLC. · 35 Tedwall Ct · Greer, SC · 29650-4791

Is this the product you have?

  • AIDBAGs are first aid kits of convenience composed of individually labeled devices designed as specialty first aid kits designed for the appropriate level provider to evaluate and treat patients at the point-of-injury / point-of-illness. The original manufacturer labeling, including single use indications, is maintained for each device, which may be used independently and at different times. These specific kits contain the McKesson TRUE Metrix PRO Professional Monitoring Blood Glucose System that were recalled
  • as part of RES 98317 with (510(k)K140100)
Lot codes and dates

2 of 2 products · 2 lot codes

Recalled lot codes, grouped by product number
Kit REFUDI-DILot
85-466100842209132089
  • 85-4661122024
85-474200842209133536
  • 85-4742123024
Amount recalled
21

Where it was sold

Why it was recalled

Stop using this product now

Kits contain the McKesson TRUE Metrix PRO Professional Monitoring Blood Glucose System that has been recalled for an issue with the software design of the E-5 Error Code where the meter displays an E-5 error code for a very high blood glucose event (> 600 mg/dL) or when there is a test strip error, which can lead to operator confusion when evaluating patients and a delay in appropriate treatment. This is a downstream recall of Trividia Health Class I, Recall Event ID 98317.

Using this product could cause serious injury, illness or death — stop using it now.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new North American Rescue LLC. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 5:41 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.