Reported Jun 6, 2018 · Terminated
Carina Ventilator, ventilator. Catalog no. 5704110. Long-term ventilator for treatment in hospitals by a qualified…
Draeger Medical, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Draeger Medical, Inc. · 3135 Quarry Rd · Telford, PA · 18969-1042
Is this the product you have?
- Carina Ventilator, ventilator. Catalog no. 5704110. Long-term ventilator for treatment in hospitals by a qualified medical personnel.
- Lot codes and dates
- Serial nos.
ASKM-0065ASKM-0066ASKM-0106ASKM-0134ASKM-0137ASKM-0150ASKM-0151ASKN-0009
- Serial nos.
- Amount recalled
- 8 units
Where it was sold
Why it was recalled
The valve assembly material may become corroded due to insufficient cleaning by a supplier. There is a risk of the breathing gas becoming contaminated in the affected devices.
This product could cause temporary or reversible health problems — stop using it.