Reported Jul 17, 2019 · Terminated

Radrex-i X-Ray System (DRAD-3000E) TFP-4336W (Wireless FPD) Software version: v5.00ER005 Product Usage: The RADREX-I is…

Canon Medical System, USA, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Canon Medical System, USA, INC. · 2441 Michelle Dr · Tustin, CA · 92780-7047

Is this the product you have?

  • Radrex-i X-Ray System (DRAD-3000E) TFP-4336W (Wireless
  • FPD) Software versionv5.00ER005
  • Product UsageThe RADREX-I is a general purpose x-ray system that employs Solid State Imager(s), SSXI, which converts x-rays directly into electrical signals which can, after appropriate processing be displayed on LCD monitors or printed to a medical grade image printer. The system console is PC based devise that allows for worklist management, image storage, image processing, image exporting and image printing. The system may be equipped with a table and/or vertical wall unit, is configurable with up to two x-ray tubes, and has an auto switching function. This system is intended for use in conjunction with the ceiling-suspended tube support, high voltage generator, and bucky stand or bucky table incorporating a fixed or detachable (portable) flat panel detector for radiography of the head, chest, abdomen, spine, neck and limbs. The system is used for image acquisition, image display and transmission/output or images to external devices
Lot codes and dates

48 of 48 products

Products covered by this recall
#Product as published
1No 1./U8A1312001/JANUARY 2013
2No.2/U8A1312002/JANUARY 2013
3No.3/U8A1342003/APRIL 2013
4No.4/U8A1362004 /JUNE 2013
5No. 5 U8B13X2006/OCTOBER 2013
6No. 6 U8B13Z2008/DECEMBER 2013
7No. 7 U8B1412009/JANUARY 2014
8No. 8 U8B1432010 /MARCH 2014
9No. 9 U8B1442011/APRIL 2014
10No.10 U8B1462013/JUNE 2014
11No.11 U8B1472014/JULY 2014
12No.12 U8B1482015/AUGUST 2014
13No.13 U8B14Y2016/NOVEMBER 2014
14No.14 U8B14Z2017/DECEMBER 2014
15No.15 U8B14Z2018/DECEMBER 2014
16No.16 U8B14Z2019/DECEMBER 2014
17No.17 U8B1512020/JANUARY 2015
18No.18 U8B1522021/FEBRUARY 2015
19No.19 U8B1522022/FEBRUARY 2015
20No.20 U8B1522023/FEBRUARY 2015
Amount recalled
46

Where it was sold

Why it was recalled

software malfunction; It was found when a user performs radiography using the wireless flat panel detector (FPD), a message window displays on the monitor stating image transmission was not completed and there was no image available. It also showed the "OK" button to reacquire image data from the FPD, and the "Cancel" button to cancel the acquisition . When the user selects the "OK" button, the same message window appears. This prompted the user to repeat the same operation several times and finally select the "Cancel" button to quit the reacquistion mode.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Canon Medical System, USA, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Jul 20, 2020 · Data as of Sep 19, 5:37 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.