Reported Jul 2, 2025 · Ongoing

Extended Tip Applicator 8CM, Box of 1.

Integra LifeSciences Corp. (NeuroSciences)

High riskUsing this product could cause serious injury, illness or death — stop using it now.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Integra LifeSciences Corp. (NeuroSciences) · 1100 Campus Rd · Princeton, NJ · 08540-6650

Is this the product you have?

    • Extended Tip Applicator 8CM
    • Box of 1
Lot codes and dates
  • Model/Catalog Number
    • 201108
    • UDI 10381780000167
    • All Lots
Amount recalled
212 units

This recall covers 4 products

The agency filed one notice per product. You are reading one of them; the other 3 are below.

Where it was sold

Why it was recalled

Stop using this product now

Devices contain potentially out-of-specification levels of endotoxins and lack of sterility assurance.

Using this product could cause serious injury, illness or death — stop using it now.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Integra LifeSciences Corp. (NeuroSciences) recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 7:40 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.