Reported Jun 18, 2025 · Ongoing
BD Texium Needle-Free Syringe: 3 mL, REF: MY8003; 5 mL, REF: MY8005; 10 mL, REF: MY8010; 20 mL, REF: MY8020; 30 mL,…
Bd Switzerland Sarl
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: BD SWITZERLAND SARL · Terre Bonne Park A4 · Eysins · N/A
Is this the product you have?
- BD Texium Needle-Free Syringe3 mL
- REF
MY80035 mL
- REF
MY800510 mL
- REF
MY801020 mL
- REF
MY802030 mL
- REF
MY803060 mL
- REFMY8060
- Lot codes and dates
6 of 6 products · 238 lot codes
Recalled lot codes, grouped by product number UDI-DI Lot (Expiration) Lot codes and dates MY80031088540322165123125144December 52026)24015468January 202027)24025611February 52027)24026390February 272027)24036107March 252027)24045647April 242027)24055859May 272027)24095513September 242027)92217803June 272025)92225801September 152025)92231303November 92025)92300201January 22026)92308702March 282026)92309501April 52026)92313601May 162026)92315701June 62026)
MY80051088540322166823125145December 52026)24015467January 202027)24026387February 272027)24045646April 242027)24055858May 272027)24075164July 262027)24095512September 242027)92215201June 12025)92220802July 272025)92225502September 122025)92315102May 312026)
MY80101088540322167524015312January 152027)24025610February 52027)24045645April 242027)24055857May 272027)24066953June 252027)24075165July 262027)24085486August 262027)24095511September 242027)24105167October 252027)92218601July 52025)92222902August 172025)92223601August 242025)92229201October 192025)92230701November 32025)92231301November 92025)92231302November 92025)92307402March 152026)92311601April 262026)92315101May 312026)
MY80201088540322168223115069November 32026)24015311January 152027)24025935February 72027)24035566March 52027)24036106March 252027)24045644April 242027)24055856May 272027)24066952June 252027)24075163July 262027)24085485August 262027)24095083September 92027)24105196October 302027)92216401June 132025)92217801June 272025)92218001June 292025)92219202July 112025)92231901November 152025)92300205January 22026)92300902January 92026)92309503April 52026)92310101April 112026)92313002May 102026)92313602May 162026)
MY80301088540323052323115070November 32026)23125143December 52026)24015264January 122027)24015644January 272027)24026389February 272027)24036105March 252027)24045643April 242027)24055855May 272027)24075162July 262027)24085484August 262027)24095509September 242027)24115110November 52027)92225501September 122025)92233602December 22025)92307403March 152026)92308701March 282026)92309502April 52026)92315702June 62026)92316402June 132026)24066951June 252027)
MY80601088540322169923115496November 282026)23125393December 182026)24015102January 82027)24026388February 272027)24036104March 252027)24045366April 232027)24055854May 272027)24075166July 262027)24095596September 242027)24115111November 142027)92216402June 132025)92222002August 82025)92222901August 172025)92224303August 312025)92224301August 312025)92226502September 222025)92228501October 122025)92230702November 32025)92230703November 32025)92230704November 32025)92300202January 22026)92300203January 22026)92306703March 82026)92306701March 82026)92306702March 82026)92311602April 262026)92312302May 32026)92313001May 102026)92316601June 152026)92316602June 152026)
- Amount recalled
- 1,265,400
This recall covers 5 products
The agency filed one notice per product. You are reading one of them; the other 4 are below.
Where it was sold
Why it was recalled
Sterile, single use closed system drug transfer devices, standalone or on infusion sets or bonded syringes, may leak at the interface between the closed male luer component and the needle-free connector, which may cause procedure delay, treatment interruptions, exposure to hazardous drugs, and/or underdosing of medication.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Bd Switzerland Sarl recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.