Reported Jul 12, 2023 · Ongoing

Quidel Triage Cardiac Panel: REF: 97000HS, 97021HS (Troponin I), 97022HS (CK-MB and Troponin I), 97000HSEU,…

Quidel Cardiovascular Inc.

High riskUsing this product could cause serious injury, illness or death — stop using it now.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: QUIDEL CARDIOVASCULAR INC · 9975 Summers Ridge Rd · San Diego, CA · 92121-2997

Is this the product you have?

  • Quidel Triage Cardiac Panel
    • REF: 97000HS
    • 97021HS (Troponin I)
    • 97022HS (CK-MB and Troponin I)
    • 97000HSEU
    • 97000HSEUJP
    • 97000HZ01
    • 97000QIL. Lots T13664 - T14045
Lot codes and dates
  • REF/UDI-DI/Lots(Expiration)97000HS/30014613337911/T13666N(2023-06-26), T13667N(2023-07-03), T13669N(2023-07-08), T13706N(2023-07-21), T13826N(2023-09-03), T13944N(2023-10-14), T13948N(2023-10-20), T13949N(2023-10-22), T14019N(2023-11-05), T14020N(2023-11-05), T14023N(2023-11-11); 97021HS/30014613337935/T13665RN(2023-06-25), T13707RN(2023-06-28), T13709RN(2023-08-11), T13827RN(2023-09-04), T13829RN(2023-09-18), T13942RN(2023-10-09), T13950RN(2023-10-23); 97022HS/30014613337942/T13664RN(2023-06-20), T13708RN(2023-08-03), T13943RN(2023-10-10); 97000HSEU/30014613337928/T13668RBN(2023-07-05), T13705RN(2023-07-14), T13765RBN(2023-08-26), T13825RBN(2023-08-27), T13828RBN(2023-09-18), T13946RBN(2023-10-16); 97000HSEUJP/T13668RN(2023-07-05), T13765RN(2023-08-26), T13828RN(2023-09-18), T13946RN(2023-10-16); 97000HZ01/T13763(2023-08-19), T13764(2023-08-21), T13830(2023-09-24), T13831(2023-09-26), T13940(2023-10-01), T13941(2023-10-03), T13947(2023-10-21), T13951(2023-10-27), T14021(2023-11-06), T14022(2023-11-10), T14024(2023-11-15), T14025(2023-11-19), T14041(2023-12-13), T14042(2023-12-18), T14043(2023-12-19), T14044(2023-12-20), T14045(2023-12-24); 97000QIL/T13825RNQ(2023-08-27). 97000HS/97021HS/97022HS:Any of the lots listed may be assigned to any of the item numbers listed. T13831N, T13831RBN, T13831RN(2023-09-26); T14021N, T14021RBN, T14021RN(2023-11-06); T14022N, T14022RBN, T14022RN(2023-11-10); T14024N, T14024RNB, T14024RN(2023-11-15); T14025N, T14025RBN, T14025RN(2023-11-19); T14041N, T14041RBN, T14041RN(2023-12-13); T14042N, T14042RBN, T14042RN(2023-12-18); T14043N, T14043RBN, T14043RN(2023-12-19); T14044N, T14044RBN, T14044RN(2023-12-20); T14045N, T14045RBN, T14045RN(2023-12-24). 97000HS/97000HSEU/97000QIL/97021HS: Any of the lots listed may be assigned to any of the item numbers listed. T13831N, T13831RBN, T13831RN, T13831NQ, T13831RBNQ, T13831RNQ(2023-09-26); T14021N, T14021RBN, T14021RN, T14021NQ, T14021RBNQ, T14021RNQ(2023-11-06); T14022N, T14022RBN, T14022RN, T14022NQ, T14022RBNQ, T14022RNQ(2023-11-10); T14024N, T14024RNB, T14024RN, T14024NQ, T14024RNBQ, T14024RNQ(2023-11-15); T14025N, T14025RBN, T14025RN, T14025NQ, T14025RBNQ, T14025RNQ(2023-11-19); T14041N, T14041RBN, T14041RN, T14041NQ, T14041RBNQ, T14041RNQ(2023-12-13); T14042N, T14042RBN, T14042RN, T14042NQ, T14042RBNQ, T14042RNQ(2023-12-18); T14043N, T14043RBN, T14043RN, T14043NQ, T14043RBNQ, T14043RNQ(2023-12-19); T14044N, T14044RBN, T14044RN, T14044NQ, T14044RBNQ, T14044RNQ(2023-12-20); T14045N, T14045RBN, T14045RN, T14045NQ, T14045RBNQ, T14045RNQ(2023-12-24). 97000HSEU/97000HSEUJP: Any of the lots listed may be assigned to any of the item numbers listed. T13831N, T13831RBN, T13831RN(2023-09-26); T14021N, T14021RBN, T14021RN(2023-11-06); T14022N, T14022RBN, T14022RN(2023-11-10); T14024N, T14024RNB, T14024RN(2023-11-15); T14025N, T14025RBN, T14025RN(2023-11-19); T14041N, T14041RBN, T14041RN(2023-12-13); T14042N, T14042RBN, T14042RN(2023-12-18); T14043N, T14043RBN, T14043RN(2023-12-19); T14044N, T14044RBN, T14044RN(2023-12-20); T14045N, T14045RBN, T14045RN(2023-12-24)
Amount recalled
16,194

Where it was sold

Why it was recalled

Stop using this product now

Raw material issue may cause cardiac panel products to have an average of 30% troponin signal reduction across measurement range 0.05 ng/mL to 30 ng/mL, QC testing won't detect anomaly, so an alternate method is recommended, if unavailable, use recommendations to mitigate potential patient impact, which could be missed/delayed myocardial infarction diagnosis.

Using this product could cause serious injury, illness or death — stop using it now.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Quidel Cardiovascular Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 6:23 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.