Reported May 20, 2020 · Terminated
Dover 100% Silicone Premium Urine Meter Tray 400mL, 16 Fr/Ch (5.3 mm), 5 mL Foley Catheter Pre-connected to 2000 mL…
Covidien LLC
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Covidien Llc · 15 Hampshire St · Mansfield, MA · 02048-1113
Is this the product you have?
- Dover 100% Silicone Premium Urine Meter Tray 400mL, 16 Fr/Ch (5.3 mm), 5 mL Foley Catheter Pre-connected to 2000 mL Drainage Bag with Temperature Sensor, Sterile
- Product CodeP4P16TS -
- Product Usagefor Drainage of the Urinary bladder and simultaneous monitoring of temperature. Prescription use only
- Lot codes and dates
- Lot Numbers
19093086641909320964 1933802964
- Lot Numbers
- Amount recalled
- 8162 units
This recall covers 11 products
The agency filed one notice per product. You are reading one of them; the other 10 are below.
Where it was sold
- Nationwide
Why it was recalled
Temperature sensing catheters failed to meet the established calibrated temperature sensing tolerance of +/- 0.2¿C. Failure to correctly measure patient body temperatures could lead to an incorrect diagnosis and insufficient treatment
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Covidien LLC recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.