Reported May 6, 2026 · Ongoing
iLet Bionic Pancreas, REF: BB1001
Beta Bionics, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Beta Bionics, Inc. · 11 Hughes · Irvine, CA · 92618-1902
Is this the product you have?
- iLet Bionic Pancreas
- REFBB1001
- Lot codes and dates
Software versions 1.4.3and 1.4.2
- Amount recalled
- 1,080
Where it was sold
Why it was recalled
insulin pump software versions 1.4.2, and 1.4.3 issue with Dexcom Continuous Glucose Monitoring (CGM) G7 sensor may result in delayed CGM reading then switch to non-CGM dosing, leading to only basal/requested meal doses provided, no correction doses/insulin reductions/suspensions for high/low glucose unless blood glucose entered for sensor life or replacement. Potential health risks to users include severe hyperglycemia with and without DKA and severe hypoglycemia.
This product could cause temporary or reversible health problems — stop using it.