Reported Jun 29, 2016 · Terminated

Biograph 16 TruePoint TV, Material Number 10249556 The Siemens Biograph TruePoint systems are combined X-Ray Computed…

Siemens Medical Solutions USA Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Siemens Medical Solutions USA Inc. · 810 Innovation Dr · Knoxville, TN · 37932-2562

Is this the product you have?

  • Biograph 16 TruePoint TV, Material Number 10249556 The Siemens Biograph TruePoint systems are combined X-Ray Computed Tomography (CT) and Positron Emission Tomography (PET) scanners that provide registration and fusion of high resolution physiologic and anatomic information.
Lot codes and dates
  • Serial Numbers
    • 72405
    • 72408
    • 72409
    • 72411
    • 72412
    • 72413
    • 72415
    • 72416
    • 72423
    • 72430
    • 72438
    • 72439
    • 72446
    • 72447
    • 72449
    • 72451
    • 72452
    • 72454
    • 72456
    • 72458
    • 72462
    • 72463
    • 72469
    • 72473
    • 72478
    • 72482
    • 72483
    • 72488
    • 72490
    • 72500
    • 72506
    • 72507
    • 72508
    • 72510
    • 72516
    • 72536
Amount recalled
36 units

This recall covers 33 products

The agency filed one notice per product. You are reading one of them; the other 32 are below.

20 more are in the archive.

Where it was sold

  • Nationwide

Why it was recalled

Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Siemens Medical Solutions USA Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated May 8, 2017 · Data as of Sep 19, 6:00 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.