Reported Jul 25, 2012 · Terminated

OEC 9900 Elite, OEC 9900 Elite MD Motorized C-arm System, OEC 9900 Elite NAV used as an aid to locate anatomical…

GE OEC Medical Systems, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: GE OEC Medical Systems, Inc · 384 Wright Brothers Dr · Salt Lake City, UT · 84116-2862

Is this the product you have?

  • OEC 9900 Elite, OEC 9900 Elite MD Motorized C-arm System, OEC 9900 Elite NAV used as an aid to locate anatomical structures during open or percutaneous surgical procedures. The system is an aid to locate anatomical structures during open or percutaneous surgical procedures. It is indicated for use in medical conditions that may benefit from the use of stereotactic surgery technique. The system provides a reference to rigid anatomical structures such as sinus, skull, long bone, or vertebra, which are visible on medical images such as CT, MRI, or X-ray.
Lot codes and dates
  • Part numbers00-887200-01; 00-887200-03; 00-887201-01; 00-887201-03; 00-887202-01; 00-887202-03; 00-887204-03; 00-887207-03; 00-887206-03. Along with all associated serial numbers
Amount recalled
726 units

This recall covers 3 products

The agency filed one notice per product. You are reading one of them; the other 2 are below.

Why it was recalled

Please be aware that this is not a new recall. The firm has taken action; but, due to administrative issues this recall is now being classified by the Agency. GE OEC recalled imaging devices OEC 9900, OEC 9800, and OEC 8800 as a result of an FDA inspection identifying that the vertical lift column power supply in the mainframe C-arm is defective and subject to early life failure.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new GE OEC Medical Systems, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Jan 28, 2013 · Data as of Sep 19, 6:11 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.