Reported Jul 16, 2014 · Terminated
SmartSite Low Sorbing Infusion Set, Model Codes: 10014855A, 10015861A, 10061661A 14035457 14035485 14046894 14035458
CareFusion 303, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: CareFusion 303, Inc. · 10020 Pacific Mesa Blvd · San Diego, CA · 92121-4386
Is this the product you have?
- SmartSite Low Sorbing Infusion Set
- Model Codes10014855A, 10015861A, 10061661A 14035457 14035485 14046894 14035458
- Lot codes and dates
- Lot No. 14035485 14046894 14035458 14035457
- Amount recalled
- 5,600 units
Where it was sold
Why it was recalled
CareFusion is recalling the SmartSite Low Sorbing Infusion Set because they were assembled with a drip chamber that consist of a 15 micron filter. The Directions For Use did not indicate that the drip chamber has a 15 micron filter and it does not provide the instruction that the set should not be used with blood administration due to the potential for occluding the 15 micron filter in the drip c
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new CareFusion 303, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.