Reported Aug 28, 2013 · Terminated
00597101101 Provisional CR PROV FRM COMP SIZE AML, Rx, Sterile; 00597101102 Provisional CR PROV FRM COMP SIZE AMR, Rx,…
Zimmer, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Zimmer, Inc. · 345 E Main St · Warsaw, IN · 46580-2746
Is this the product you have?
- 00597101101 Provisional CR PROV FRM COMP SIZE AML, Rx, Sterile; 00597101102 Provisional CR PROV FRM COMP SIZE AMR, Rx, Sterile; 00597101201 Provisional CR PROV FRM COMP SIZE BML, Rx, Sterile; 00597101202 Provisional CR PROV FRM COMP SIZE BMR, Rx, Sterile; 00597101303 Provisional CR PROV FRM COMP SIZE CML, Rx, Sterile; 00597101304 Provisional CR PROV FRM COMP SIZE CMR, Rx, Sterile; 00597101403 Provisional CR PROV FRM COMP SIZE DML, Rx, Sterile; 00597101404 Provisional CR PROV FRM COMP SIZE DMR, Rx, Sterile; 00597101503 Provisional CR PROV FRM COMP SIZE EML, Rx, Sterile; 00597101504 Provisional CR PROV FRM COMP SIZE EMR, Rx, Sterile; Used in total knee arthroplasty and indicated for patients with severe pain.
- Lot codes and dates
- all codes
- Amount recalled
- 192,355 all devices
This recall covers 48 products
The agency filed one notice per product. You are reading one of them; the other 47 are below.
35 more are in the archive.
Where it was sold
- Nationwide
Why it was recalled
Surgeons are implanting components of the NexGen Cruciate Retaining (CR) Complete Knee System in combinations that are not approved (micro with standard). Includes CR Micro Femurs, Articular Surface, Patellae, instruments and provisionals combined with standard components. A March 2012 recall reiterated the proper combination labeling, but the mis match problem continues. Zimmer is removing the d
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Zimmer, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.