Reported May 10, 2017 · Terminated

Sysmex, Automated Hematology Analyzer XN Series, Sysmex Corporation, 1-5-1 Wakinohama-Kaigandori, Chuo-ku, Kobe…

Sysmex America Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Sysmex America Inc · 577 Aptakisic Rd · Lincolnshire, IL · 60069-4325

Is this the product you have?

  • Sysmex, Automated Hematology Analyzer XN Series, Sysmex Corporation, 1-5-1 Wakinohama-Kaigandori, Chuo-ku, Kobe 651-0073 Japan
Lot codes and dates
  • **Model: XN-10
  • Product NumberAP795756
  • Serial Numbers
    • 11001  29656
    • 29668-29786
    • 29788-29798
    • 29800-29801
    • 29803
    • 29804
    • 29807
    • 29812
    • 29816-29867 **Model: XN-11
  • Product NumberCS297931
  • Serial Numbers
    • A1001-A1124
    • A1127
    • A1130
    • A1135-A1140
    • A1142-A1144
    • A1146-A1149
    • A1151
    • A1153-A1159
    • A1161
    • A1163
    • A1171
Amount recalled
3,682 units

Where it was sold

  • Nationwide

Why it was recalled

Reported customer complaints of automated hematology analyzers catching fire due to misalignment of an internal protective sheet that shields electronic components from salty reagents. Build-up of salt deposits on unprotected electronic components may combust due to heat generated by electronics when instrument cycles. Completed CAPA identified the root cause as supplier providing defective component (sensor ready valve) and the misalignment of the electronic protective shield during manufacturing of the device.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Sysmex America Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Aug 22, 2019 · Data as of Sep 19, 4:47 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.