HeartStart MRx Monitor/Defibrillator, Model No. M3535A (861288), M3536A (861289) Product Automated external…
Philips Electronics North America Corporation
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Philips Electronics North America Corporation · 3000 Minuteman Rd · Andover, MA · 01810-1032
Is this the product you have?
- HeartStart MRx Monitor/Defibrillator, Model No. M3535A (861288), M3536A (861289) Product Automated external defibrillators (non-wearable)
- Product UsageThe HeartStart MRx is for use for the termination of ventricular tachycardia and ventricular fibrillation. The device is for use by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac support, or defibrillation. It must be used by or on the order of a physician
- Lot codes and dates
US00588464US00588657US00589878US00591080US00593861
- Amount recalled
- 5
Where it was sold
Why it was recalled
The MRx monitor/defibrillators could fail to charge because the therapy printed circuit board may have been loaded with an incorrect electronic component. Failure to charge could potentially cause therapy to be interrupted or delayed.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Philips Electronics North America Corporation recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.