Reported May 27, 2020 · Terminated

Professional Urine Drug Control, Item code 88010 - Product Usage: The Alere Professional Cup Urine Drug Controls are…

Cardinal Health Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Cardinal Health Inc. · 7000 Cardinal Pl · Dublin, OH · 43017-1091

Is this the product you have?

  • Professional Urine Drug Control
  • Item code 88010 - Product UsageThe Alere Professional Cup Urine Drug Controls are compatible with all quantitative and qualitative drug detection procedures which are sufficiently sensitive to detect the control constituents
Lot codes and dates
  • Lot KN00146
Amount recalled
N/A

This recall covers 3 products

The agency filed one notice per product. You are reading one of them; the other 2 are below.

Where it was sold

Why it was recalled

Three products requiring refrigerated storage conditions were incorrectly stored outside of approved temperature ranges. Using these temperature abused products may result in misdiagnosis of the patient.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Cardinal Health Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Jul 23, 2020 · Data as of Sep 19, 4:49 PM UTC